Weld Probe Micro-Force Calibrator
WP9000

FORCE RANGES TO 250 LBS.

SUITED TO MICROJOINING

NIST TRACEABLE ACCURACY 0.25

RELIABLE, REPEATABLE RESULTS

WELD THROUGH IT

The WeldProbe is a rugged, hand-held instrument for accurately calibrating weld electrode force on table-tip and micro joining resistance welding heads. This device helps ensure compliance with ISO9000 or QS9000 quality requirements by taking the guesswork out of setup. Measuring accurate force levels at the electrode tips helps eliminate innaccuracies of line-pressure readings, springs, lever arms and mechanical friction and wear.

The WeldProbe is designed for long life in the harsh welding evvironment by using standard fully-welded load cell construction and NEMA 4 electronic packaging. The easy to use design includes a simple push-button format and a large LCD display that provides both continuous and peak (max) force values in keypad selectable engineering units.

High-speed digital electronics sample force readings during the squeeze-weld-hold sequence and capture the peak force for display. A simple, one-button CLEAR prepares the instrument for the next reading. Two standard 9 volt alkaline batteries provide 80 hours of continuous use. Battery life is maximized with a programmable power-off feature.

SPECIFICATIONS

PERFORMANCE

Capacities

25,100,250 lbs.

Overload

150%

Accuracy

0.25%

Output

4 digit LCD display

Sampling Rate

19.2 kHz internal;24Hz display update

Battery Life
(continuous)

80 hours with two 9 volt alkaline batteries
(includes programmable auto-off features)

Display

Selectable engineering units (lbs.,Kg,N)

ENVIRONMENTAL

Package

All welded load cell, NEMA 4 electronics

Operating
Temperature

30 deq F. to 140 deg F.

Weld Current

Survives accidental current discharge

Approvals

CE rated for EMC environments

How to Order

Model WP9000
             25 lbs.         BL
           100 lbs.         BR
           250 lbs.         CN

Head Configuration
             Indents         Standard
             Projection Button 72a

 

Operating Sequence

1. Depress ON/OFF button to power up unit.
2. Press and hold ZERO button for zero display
3. Insert probe and take force reading
4. Toggle MIN/MAX button to read MAX (peak) force value
5. Press CLEAR to reset the MAX value in preparation for next reading